| Reactivity | Universal |
| Detection Range | 312.5-20000 pg/mL |
| Sensitivity | 92.5 pg/mL |
This ELISA kit uses the Compitive-ELISA principle. The micro-ELISA plate provided in this kit has been pre-coated with an antibody specific to cAMP. Samples (or Standards) are added to the micro-ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for cAMP and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain cAMP, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of cAMP. You can calculate the concentration of cAMP in the samples by comparing the OD of the samples to the standard curve.
| Catalog Number | LAB13344 |
| Reactivity | Universal |
| Detection Range | 312.5-20000 pg/mL |
| Sensitivity | 92.5 pg/mL |
Product name: | Lablisa® cAMP (Cyclic Adenosine Monophosphate) ELISA Kit |
| Reactivity: | Universal |
| Alternative Names: | c-AMP; 3'-5'-Cyclic Adenosine Monophosphate; Adenosine Cyclophosphate |
| Assay Type: | Compitive |
| Sensitivity: | 92.5 pg/mL |
| Standard: | 20000 pg/mL |
| Range: | 312.5-20000 pg/mL |
| Sample Type: | Serum, Plasma and Tissue homogenate and other biological samples;Sample volume:25μL |
| Assay Length: | 3H |
| Research Area: | Signal transduction;Metabolic pathway; |
Standard curve
| Concentration (pg/mL) | OF |
|---|---|
| 20000 | 0.195 |
| 10000 | 0.331 |
| 5000 | 0.546 |
| 2500 | 0.88 |
| 1250 | 1.308 |
| 625 | 1.7 |
| 312.5 | 2.076 |
| 0 | 2.516 |
Precision
Intra-assay Precision (Precision within an assay) :CV%<8% Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) :CV%<10% Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant cAMP and the recovery rates were calculated by comparing the measured value to the expected amount of cAMP in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum (n=8) | 80-92% | 86% |
| EDTA plasma (n=8) | 90-105% | 97% |
| Cell culture media (n=8) | 88-102% | 95% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of cAMP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum (n=8) | 85-94% | 87-99% | 85-101% | 80-90% |
| EDTA plasma (n=8) | 86-97% | 86-97% | 88-95% | 92-101% |
| Cell culture media (n=8) | 87-98% | 85-92% | 80-96% | 89-100% |
Manuals on the web are for your reference only, specifications are subject to the delivery manuals.
Available Soon...
Share your experience with this product to help other researchers.
Login to Submit a ReviewNo reviews yet. Be the first to share your experience!