ELISA Kits

Lablisa® F-E3(Free Estriol) ELISA Kit

Catalog No.: LAB13324
Competitive
ReactivityUniversal
Detection Range0.63-40 ng/mL
Sensitivity0.218 ng/mL
₹32,550.00
This product is currently out of stock
Dispatch: Usually dispatched within 5–10 working days

Description

This ELISA kit uses the Competitive-ELISA principle. The micro-ELISA plate provided in this kit has been pre-coated with F-E3. During the reaction, F-E3 in the sample or standard competes with a fixed amount of F-E3 on the solid phase supporter for sites on the Biotinylated Detection Ab specific to F-E3. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The concentration of F-E3 in tested samples can be calculated by comparing the OD of the samples to the standard curve.

Technical Specifications

Catalog NumberLAB13324
ReactivityUniversal
Detection Range0.63-40 ng/mL
Sensitivity0.218 ng/mL

Product Overview

Product name:

Lablisa® F-E3(Free Estriol) ELISA Kit
Reactivity:Universal
Alternative Names:F-E3; Free Estriol
Assay Type:Competitive
Sensitivity:0.218 ng/mL
Standard:40 ng/mL
Range:0.63-40 ng/mL
Sample Type:Serum, Plasma and Tissue homogenate and  other biological samples;Sample volume:25μL
Assay Length:1.5H
Research Area:Endocrinology;Reproductive science;Hormone metabolism;

Standard curve

Concentration (ng/mL)OD
400.232
200.333
100.608
50.894
2.51.199
1.251.61
0.631.909
02.702

Precision

Intra-assay Precision (Precision within an assay) :CV%<8% Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) :CV%<10% Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant F-E3 and the recovery rates were calculated by comparing the measured value to the expected amount of F-E3 in samples.

MatrixRecovery rangeAverage
serum (n=8)86-96%91%
EDTA plasma (n=8)91-99%94%
Cell culture media (n=8)93-99%96%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F-E3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum (n=8)95-103%90-100%90-98%93-101%
EDTA plasma (n=8)94-102%93-99%90-97%90-97%
Cell culture media (n=8)91-100%97-104%98-104%91-99%

Manuals on the web are for your reference only, specifications are subject to the delivery manuals.

Product Images

Lablisa® F-E3(Free Estriol) ELISA Kit

Documents

Datasheet
Download

Protocol

For detailed protocol, assay guidance, and technical support, please contact our Customer Support team at