ELISA Kits

Lablisa® Human ACTH (Adrenocorticotropic Hormone) ELISA Kit

Catalog No.: LAB9299
Competitive
ReactivityHuman
Detection Range15.63-1000pg/mL
Sensitivity9.38pg/mL
₹32,550.00
Dispatch: Usually dispatched within 5–10 working days

Description

This ELISA kit uses the Competitive-ELISA principle. The micro-ELISA plate provided in this kit has been pre-coated with Human ACTH. During the reaction, Human ACTH in the sample or standard competes with a fixed amount of Human ACTH on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human ACTH. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The concentration of Human ACTH in tested samples can be calculated by comparing the OD of the samples to the standard curve.

Technical Specifications

Catalog NumberLAB9299
ReactivityHuman
Detection Range15.63-1000pg/mL
Sensitivity9.38pg/mL

Product Overview

Product name:

Lablisa® Human ACTH (Adrenocorticotropic Hormone) ELISA Kit
Reactivity:Human
Alternative Names:Corticotropin, Adrenocorticotropin
Assay Type:Competitive
Sensitivity:9.38 pg/mL
Standard:1000 pg/mL
Range:15.63-1000 pg/mL
Sample Type:Serum, Plasma and Tissue homogenate and  other biological samples;Sample volume:25μL
Assay Length:3H
Research Area:Signal transduction;

Standard curve

Concentration (pg/mL)OD
10000.238
5000.402
2500.653
1250.841
62.51.193
31.251.508
15.631.987
02.727

Precision

Intra-assay Precision (Precision within an assay) :CV%<8% Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) :CV%<10% Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant Human ACTH and the recovery rates were calculated by comparing the measured value to the expected amount of Human ACTH in samples.

MatrixRecovery rangeAverage
serum(n=8)92-100%96%
EDTA plasma(n=8)89-95%92%
Cell culture media (n=8)88-97%93%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Human ACTH and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=8)87-95%93-101%89-98%96-104%
EDTA plasma(n=8)96-104%97-103%95-103%99-107%
Cell culture media (n=8)96-105%89-97%89-96%93-102%

Manuals on the web are for your reference only, specifications are subject to the delivery manuals.

Product Images

Lablisa® Human ACTH (Adrenocorticotropic Hormone) ELISA Kit

Documents

Datasheet
Download

Protocol

For detailed protocol, assay guidance, and technical support, please contact our Customer Support team at
support@labrecon.in

Publications

Available Soon...

Customer Reviews

0.0
  • Star
  • Star
  • Star
  • Star
  • Star
Based on 0 reviews

Share your experience with this product to help other researchers.

Login to Submit a Review

No reviews yet. Be the first to share your experience!

Feedback & Reviews