| Reactivity | Human |
| Detection Range | 15.63-1000ng/mL |
| Sensitivity | 9.38ng/mL |
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human DHEA-S. During the reaction, Human DHEA-S in the sample or standard competes with a fixed amount of Human DHEA-S on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human DHEA-S. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The concentration of Human DHEA-S in tested samples can be calculated by comparing the OD of the samples to the standard curve.
| Catalog Number | 15.63-1000ng/mL |
| Reactivity | Human |
| Detection Range | 15.63-1000ng/mL |
| Sensitivity | 9.38ng/mL |
Product name: | Lablisa® Human DHEA-S (Dehydroepiandrosterone Sulfate) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | Androstenolone sulfate; Prasterone sulfate; Androst-5-en-3β-ol-17-one 3β-sulfate |
| Assay Type: | Competitive |
| Sensitivity: | 9.38ng/mL |
| Standard: | 1000ng/mL |
| Range: | 15.63-1000ng/mL |
| Sample Type: | Serum, Plasma and Tissue homogenate and other biological samples;Sample volume:25μL |
| Assay Length: | 1.5H |
| Research Area: | Enzyme & Kinase; |
Standard curve
| Concentration (pg/mL) | OD |
|---|---|
| 1000 | 0.226 |
| 500 | 0.329 |
| 250 | 0.554 |
| 125 | 0.92 |
| 62.5 | 1.236 |
| 31.25 | 1.491 |
| 15.63 | 2.017 |
| 0 | 2.58 |
Precision
Intra-assay Precision (Precision within
an assay) :CV%<8% Three samples
of known concentration were tested twenty
times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) :CV%<10% Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Human DHEA-S and the recovery rates were calculated by comparing the measured value to the expected amount of Human DHEA-S in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=8) | 86-93% | 90% |
| EDTA plasma(n=8) | 88-97% | 93% |
| Cell culture media (n=8) | 92-102% | 97% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Human DHEA-S and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=8) | 90-98% | 90-98% | 95-101% | 90-98% |
| EDTA plasma(n=8) | 97-105% | 90-97% | 97-105% | 99-107% |
| Cell culture media (n=8) | 90-96% | 90-96% | 92-101% | 88-96% |