| Reactivity | Human |
| Detection Range | 0.16-10mIU/mL |
| Sensitivity | 0.09 mIU/mL |
This ELISA kit uses the Sandwich-ELISA principle. The micro-ELISA plate provided in this kit has been pre-coated with an antibody specific to Human LH. Samples (or Standards) are added to the micro-ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human LH and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human LH, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of Human LH. You can calculate the concentration of Human LH in the samples by comparing the OD of the samples to the standard curve.
| Catalog Number | LAB13635 |
| Reactivity | Human |
| Detection Range | 0.16-10mIU/mL |
| Sensitivity | 0.09 mIU/mL |
Product name: | Lablisa® FastTrack Human LH (Luteinizing Hormone) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | Lutropin, Lutrophin |
| Assay Type: | Sandwich |
| Sensitivity: | 0.09 mIU/mL |
| Standard: | 10 mIU/mL |
| Range: | 0.16-10mIU/mL |
| Sample Type: | Serum, Plasma and Other biological fluids;Sample Volume=50μL |
| Assay Length: | 1.5h |
| Research Area: | Signaling transduction, Neuroscience, Cancer, Developmental Biology, Metabolism |
Standard curve
| Concentration (ng/mL) | OD | Corrected |
|---|---|---|
| 10 | 3.194 | 3.193 |
| 5 | 2.963 | 2.962 |
| 2.5 | 2.414 | 2.413 |
| 1.25 | 1.491 | 1.49 |
| 0.63 | 0.704 | 0.703 |
| 0.31 | 0.349 | 0.348 |
| 0.16 | 0.171 | 0.17 |
| 0 | 0.001 | 0 |
Precision
Intra-assay Precision (Precision within an assay) :CV%<8% Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) :CV%<10% Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Human LH and the recovery rates were calculated by comparing the measured value to the expected amount of Human LH in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=8) | 80-95% | 87% |
| EDTA plasma(n=8) | 83-98% | 90% |
| Cell culture media (n=8) | 86-99% | 92% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Human LH and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=8) | 80-94% | 87-96% | 85-98% | 95-104% |
| EDTA plasma(n=8) | 85-103% | 85-103% | 82-96% | 87-105% |
| Cell culture media (n=8) | 90-99% | 88-102% | 86-92% | 81-93% |